ATI A15/76 Turbidity Monitor Part 5 – Configuration
23
O&M Manual
Rev-H (8/11)
19. AL TYPE: CONT. This routine allow definition of the operation of the alarm relay. The alarm
relay C can be programmed for either continuous or flash operation. In continuous mode, an
alarm will cause the relay coil to change state and stay in that position. In “flash” mode, the relay
coil will toggle so that anything connected through the relay will alternate from on to off. If you
program the alarm for flash mode, you then have to select a high, medium, or low rate of flash.
20. CAL CF: DISABLED This routine is used to activate an automatic sensor cleaning system
available for a different version of this product. This setting should remain Disabled.
21. CLEANING T: 15.0” This routine is associated with the Auto-Clean version of this product and
is not used in systems employing the EPA approved type of sensor design.
22. HOLDING T: 3.0’ As with #21 above, this is a routine associated with the Auto-Clean version of
this product.
23. Change A Nr.: No/Yes: This routine allows you to enter your own access code number.
Changing the access number will restrict access to the configuration menu by anyone who does
not know the number you enter. To change the number, press UP to change the display to YES
and press ENTER. Use the UP or DOWN arrow to set the number you want and then press
ENTER. When you press ENTER, the display will show "Confirm Nr.: 0". To change the access
number, the number must be entered twice. Use the UP or DOWN arrows to set the same
number again and then press ENTER. After the second entry, the new number will be stored and
must be used to gain access to the configuration menu from then on.