ATI A15/66 Residual Sulfite Monitor
10
O&M Manual
Rev-W (8/10)
To wire the chemistry module, you need to open the center compartment of the assembly. On
the left side of the enclosure is a slot for releasing the latch. Slide a small screwdriver into the slot and
push the screwdriver handle toward the side of the enclosure. The latch will snap open and you can
release the catch, allowing the front part of the chemistry module to swing open on the right hand hinge.
If you wish, you can completely remove the front section by unplugging the pump terminal block, the
upper terminal block marked “to sensor”, and disconnecting the air line from the pump. With these
disconnected, you can open the right hand latch and remove the entire front section.
Figure 6 - Chemsitry Module Connections